![]() ![]() We sought to validate the accuracy of the Doser with several commonly prescribed MDIs. In the laboratory, clinic personnel actuated an MDI with an attached Doser several times in succession on 3 consecutive days and recorded each of the 3 measures of MDI use (study 1). In study 2 clinic personnel carried an MDI and attached Doser with them for 4 weeks, actuating the MDI according to a prescribed protocol each morning and evening and again recording each of the 3 measures of daily use. In addition, during 2 weeks of study 2, a thermistor-based Nebulizer Chronolog was attached to the MDI to electronically record the date and time of each actuation. In study 3 clinic patients had both a Doser and Nebulizer Chronolog attached to their routinely used inhalers for 2 weeks and a Doser alone during a separate 2-week period. In study 1 agreement was 99% to 100% among the 3 Doser measures, and each measure agreed with actual use by self-report 97% of the time. In study 2 agreement among the 3 Doser measures of use ranged from 98% to 99%. Agreement between each of the 3 Doser measures and the Nebulizer Chronolog ranged from 90% to 93%. Agreement between each of the 3 Doser measures and actual use ranged from 96% to 97%, and the Nebulizer Chronolog agreed with actual use 93% of the time.